Trace Uranium Analyzer
Application Fields
Measuring Uranium content in fields of geological survey, nuclear fuel industry, environmental protection, industrial hygiene, food inspection, health and epidemic prevention system, etc.
Instrument Characteristics
Measurement mode: automatic measurement mode and manual measurement mode.
It can choose reasonable measuring gear intelligently according to the sample content in automatic measurement mode.
Wide range. When content is less than 6000 ng/mL, sample can be measured directly without dilution operation.
Display the result of liquid sample or solid sample according to the the original state of the sample (µg/mL or µg/g).
Import ultraviolet source, long lifetime, analyze at least 80000 uranium samples.
TFT 5 inch color LCD display.
Equipped with embedded thermal printer, the measured results can be directly printed.
Instrument can be connected PC through USB and support for remote control operation.
Specifications
The instrument meets the requirements of EJ/T 823 standard, the main performance indexes are as follows:
1. Measuring objects: Liquid sample. Samples need to be prepared and converted to liquid samples
2. Detection Limit: ≤ 0.01 ng/mL (three times standard deviation)
3. Precision: RSD ≤ 5%
4. Accuracy: MPE ±8%
5. Measuring Range: (0 – 6000) ng/mL
6. Linearity: Correlation coefficient ≥ 0.9992
7. Stability (8h): RE ≤ 7% (Uranium concentration: 2.00 ng/mL)
8. Data Print: Thermal micro printer
9. Communication Type: USB
10. Storage Capacity: 4000 sets of measuring data
11. Operating Environment: Temperature: (+10 – +40)°C Relative humidity: < 85% (+30°C)
12. Power/Power Consumption: 220V/50Hz, about 50 W
13. Dimension and Weight: (45×36×19) cm 15 kg